Loading HuntDB...

CVE-2023-25931

MEDIUM
Published 2023-03-01T18:56:30.093Z
Actions:

Expert Analysis

Professional remediation guidance

Get tailored security recommendations from our analyst team for CVE-2023-25931. We'll provide specific mitigation strategies based on your environment and risk profile.

CVSS Score

V3.1
6.4
/10
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:H
Base Score Metrics
Exploitability: N/A Impact: N/A

EPSS Score

v2025.03.14
0.001
probability
of exploitation in the wild

There is a 0.1% chance that this vulnerability will be exploited in the wild within the next 30 days.

Updated: 2025-06-25
Exploit Probability
Percentile: 0.312
Higher than 31.2% of all CVEs

Attack Vector Metrics

Attack Vector
PHYSICAL
Attack Complexity
LOW
Privileges Required
NONE
User Interaction
NONE
Scope
UNCHANGED

Impact Metrics

Confidentiality
LOW
Integrity
HIGH
Availability
HIGH

Description

Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.

Available Exploits

No exploits available for this CVE.

Related News

No news articles found for this CVE.

Affected Products

GitHub Security Advisories

Community-driven vulnerability intelligence from GitHub

⚠ Unreviewed MODERATE

GHSA-crgr-72p3-3g72

Advisory Details

Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.

CVSS Scoring

CVSS Score

5.0

CVSS Vector

CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

Advisory provided by GitHub Security Advisory Database. Published: March 1, 2023, Modified: March 9, 2023

References

Published: 2023-03-01T18:56:30.093Z
Last Modified: 2025-03-07T21:34:37.484Z
Copied to clipboard!