CVE-2024-48966
Expert Analysis
Professional remediation guidance
Get tailored security recommendations from our analyst team for CVE-2024-48966. We'll provide specific mitigation strategies based on your environment and risk profile.
CVSS Score
V3.1EPSS Score
v2025.03.14There is a 0.1% chance that this vulnerability will be exploited in the wild within the next 30 days.
Attack Vector Metrics
Impact Metrics
Description
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
Available Exploits
Related News
Affected Products
Affected Versions:
Affected Versions:
GitHub Security Advisories
Community-driven vulnerability intelligence from GitHub
Advisory Details
CVSS Scoring
CVSS Score
CVSS Vector
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
References
Advisory provided by GitHub Security Advisory Database. Published: November 15, 2024, Modified: November 15, 2024