GHSA-285j-vfp9-5w9w
GitHub Security Advisory
⚠ Unreviewed
CRITICAL
Has CVE
Advisory Details
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
Related CVEs
Key Information
9.0
/10
Dataset
Last updated: July 12, 2025 6:29 AM
Data from GitHub Advisory Database. This information is provided for research and educational purposes.