Loading HuntDB...

GHSA-2wv9-4g7m-79pc

GitHub Security Advisory

⚠ Unreviewed MODERATE Has CVE

Advisory Details

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes.

Related CVEs

Key Information

GHSA ID
GHSA-2wv9-4g7m-79pc
Published
May 13, 2022 1:38 AM
Last Modified
May 13, 2022 1:38 AM
CVSS Score
5.0 /10
Primary Ecosystem
Unknown
Primary Package
Unknown
GitHub Reviewed
✗ No

Dataset

Last updated: July 12, 2025 6:29 AM

Data from GitHub Advisory Database. This information is provided for research and educational purposes.