Loading HuntDB...

GHSA-ccgj-ph26-hf87

GitHub Security Advisory

⚠ Unreviewed MODERATE Has CVE

Advisory Details

Certain 802.11 network management messages have been determined to invoke wireless access point blacklisting security defenses when not required, which can necessitate intervention by hospital staff to reset the device and reestablish a network connection to the Wi-Fi access point. During this state, the Philips IntelliVue MX40 Version B.06.18 can either connect to an alternative access point within signal range for association to a central monitoring station, or it can remain in local monitoring mode until the device is reset by hospital staff. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.

Related CVEs

Key Information

GHSA ID
GHSA-ccgj-ph26-hf87
Published
May 13, 2022 1:36 AM
Last Modified
May 13, 2022 1:36 AM
CVSS Score
5.0 /10
Primary Ecosystem
Unknown
Primary Package
Unknown
GitHub Reviewed
✗ No

Dataset

Last updated: July 7, 2025 6:28 AM

Data from GitHub Advisory Database. This information is provided for research and educational purposes.